Temperature sensors and transducers
Pt100, Pt1000 / current 4-20mA
Temperature sensor P0120
P0120 - air temperature transmitter
outdoor, indoor use
TECHNICAL PARAMETERS | |
---|---|
Temperature measuring and operating range: |
-30 to +80°C |
Accuracy: | ±0.4°C |
Applied sensor: | Pt1000/3850ppm |
Break of sensor: | >24mA |
Short circuit of sensor: | <3.5mA |
Dimensions (W x H x D): | approximately 89 x 151 x 39,5 mm |
Protection: | IP65 |
Power: | 9 to 30 V dc (typically 24V) |
Warranty: | two years |
Temperature sensor P0132
P0132 - duct temperature transmitter
TECHNICAL PARAMETERS | |
---|---|
Temperature measuring and operating range: |
0 to +150°C |
Accuracy: |
±0.4°C up to 100°C 0,4% from reading over 100°C |
Applied sensor: | Pt1000/3850ppm |
Break of sensor: | >24mA |
Short circuit of sensor: | <3.5mA |
Dimensions (W x H x D): | approximately 89 x 220 x 39,5 mm, stem length 120mm |
Protection: | IP65 |
Power: | 9 to 30 V dc (typically 24V) |
Warranty: | two years |
Temperature transducer P6181
Temperature transducers from Pt100, Pt1000 sensor to current 4-20mA
TECHNICAL PARAMETERS | |
---|---|
Operating temperature: |
-30 to +80°C |
Break of sensor: | >24mA |
Short circuit of sensor: | <3.5mA |
Dimensions (W x H x D): | approximately 89 x 123 x 39,5 mm |
Protection: | IP65 |
Power: | 9 to 30 V dc (typically 24V) |
Warranty: | two years |
Pt100 transducer
two-wire, three-wire and two-wire with compensation loop connectionTyp |
Temperature range of 4-20mA output |
Accuracy of the input |
---|---|---|
P6181 | -100 to +200°C |
±0,3 °C up to +100°C, ±0,4°C over +100°C |
Manual for P6181 in pdf format
Pt1000 transducers, two-wire connection
Temperature transducer P41x1
Typ |
Temperature range of 4-20mA output |
Accuracy of the input |
---|---|---|
P4141 | -100 to +30°C | ±0.3°C |
P4191 | -50 to +50°C | ±0.3°C |
P4121 | -30 to +80°C | ±0.3°C |
P4151 | 0 to +35°C | ±0.2°C |
P4131 | 0 to +150°C | ±0.3°C |
P4161 | 0 to +250°C | ±0.4°C |
P4171 | 0 to +400°C | ±0.7°C |
Transducers P41x1 are directly compatible with Comet temperature probes with the Pt1000 sensor with no connector - probe marking followed by /0 .
Manual for P0120 and P41x1 in pdf format
Included is traceable calibration certificate from the manufacturer, instruction manual.
Calibration certificate with declared metrological traceability of etalons is based on requirements of EN ISO/IEC 17025 standard.
Example of calibration certificate.
Typical application wiring
Calculation of minimum power voltage Uss for proper operation:
Uss min> Uo min + Imax * Rz,
where: Uo min= 9 V
Imax... approximately 20 mA
Rz... sensing resistor (shunt)
Transmitters are directly compatible with sixteen channel Comet data acquisition system MS.