Temperature, humidity, barometric pressure sensors with serial interface RS485
- temperature
- barometric pressure
- relative humidity
- dew point temperature
- absolute humidity
- specific humidity
- mixing ratio
- specific enthalpy
Measured temperature, relative humidity, barometric pressure are converted to digital serial output with the RS485 link parameters. Transmitters are available in wall-mount, duct-mount and bar types.
Also types with T+RH probe on a cable are available. Model T3419P is designed for compressed air measurement up to 25 bars.
Transmitter circuitry is galvanically isolated from power circuitry to prevent collision in RS485 network.
Large dual line LCD for display of temperature, humidity, pressure or other calculated humidity interpretation is an advantage. Computerized design ensures maximum long term stability and fail indication. The transmitter works with ModBus RTU communication protocol or with Advantech ADAM compatible protocol. Protocol is user selectable in special configuration mode by means of the PC. Serial link enables to read actual readings and modify transmitter configuration. Instrument works always in slave mode, i.e. responds only to master device query.
Ask more information on availability of other communication protocols.
State-of-the-art capacitive polymer humidity sensor excels in mechanical stress, temperature shock, dew resistance as well as in large linearity range and low hysteresis. Transmitters are designed for use in non-aggressive environment.
APPLICATIONS - measuring of temperature, humidity, pressure at:
- building HVAC management
- building HVAC automation
- warehouses
- glasshouses
- manufacturers
- museums, archives, galleries
- air-conditioned rooms
- weather stations
COMMON TECHNICAL PARAMETERS | |
---|---|
Supported temperature units: | degrees Celsius, degrees Fahrenheit |
Temperature sensor: | RTD Pt1000-3850ppm/°C |
Operating temperature range of the case with electronics: |
-30 to +80°C, recommended to switch OFF LCD display over +70°C |
Accuracy: | ±0.4°C |
Measuring relative humidity range: | 0 to 100% |
Accuracy: | ±2.5% RH from 5 to 95% at 23°C |
Accuracy and range of dew point temperature output:
- for more details see graphs |
±1,5°C at ambient temperature T<25°C and RH>30% range -60 to +80 °C |
Accuracy and range of absolute humidity output: | ±3g/m3 at ambient temperature T < 40°C range 0 to 400 g/m3 |
Accuracy and range of specific humidity output: | ±2g/kg at ambient temperature T < 35°C range 0 to 550 g/kg |
Accuracy and range of mixing ratio output: | ±2g/kg at ambient temperature T < 35°C range 0 to 995 g/kg |
Accuracy and range of specific enthalpy output: | ±3kJ/kg at ambient temperature T < 25°C range: 0 to 995 kJ/kg |
Accuracy and range of barometric pressure output: | ± 1.3hPa at 23°C range: 600 to 1100hPa |
Supported pressure units: | hPa, kPa, mbar, mmHg, inHg, inH2O, PSI, oz/in^2 |
Temperature compensation of the humidity sensor: | from -30 to +125°C |
Power: | 9 to 30 V DC, power consumption approximately 0.5W |
Protection - temperature and humidity transmitters: |
IP65 electronics with terminals, IP40 humidity and temperature sensors |
Protection - transmitters measuring pressure: |
IP54 electronics with terminals, IP40 humidity and temperature sensors |
Filtering ability of sensor cover: | 0.025mm - filter from stainless steel mesh |
Communication protocol: |
ModBus RTU and Advantech ADAM compatible protocol
![]() |
Communication speed: | 110 to 115200 Bd |
Warranty: | two years |
TRANSMITTER MODELS | |||||
---|---|---|---|---|---|
Communication circuits are GALVANICALLY ISOLATED from power circuitry to prevent collision in RS485 network | |||||
Type | Measured value | Description | Manual in pdf format | ||
T4411![]() |
temperature |
Temperature transduser for RTD sensors Pt1000 outdoor and indoor use maximum range of temperature measurement -200 to +600°C |
![]() |
||
T0410![]() |
temperature |
Thermometer outdoor and indoor use stem length 53mm maximum range of temperature measurement -30 to +80°C |
![]() |
||
T3411![]() |
temperature humidity |
Thermometer-hygrometer outdoor and indoor use stem length 75mm maximum range of temperature measurement -30 to +80°C dimensions |
![]() |
||
T3413![]() |
temperature humidity |
Thermometer-hygrometer duct mount stem length 150mm maximum range of temperature measurement -30 to +125°C* |
![]() |
||
T3413D![]() |
temperature humidity |
Thermometer-hygrometer duct mount stem length 150mm maximum range of temperature measurement -30 to +125°C* |
![]() |
||
T3417![]() |
temperature humidity |
Thermometer-hygrometer bar type stem length 700mm maximum range of temperature measurement -30 to +125°C* |
![]() |
||
T3419![]() |
temperature humidity |
Thermometer-hygrometer T+RH probe with 1 m cable available also with probe cable lengths 2m or 4m maximum range of temperature measurement -30 to +105°C* probe including cable |
![]() |
||
New !T3419P![]() ![]() |
temperature humidity |
Thermometer-hygrometer T+RH probe from hard anodized duralumin with 1 m cable for compressed air up to 25 bars. Available also with probe cable lengths 2m or 4m. Maximum range of temperature measurement -30 to +105°C* probe including cable. |
![]() |
||
T7410![]() |
temperature humidity barometric pressure |
Thermometer-hygrometer-barometer outdoor and indoor use stem length 75mm Barometer enables to measure sea level pressure by setting of correction to altitude above sea level. maximum range of temperature measurement -30 to +80°C dimensions |
![]() |
||
T7411![]() |
temperature humidity barometric pressure |
Thermometer-hygrometer-barometer T+RH probe with 1 m cable. Available also with probe cable lengths 2m or 4m. Pressure sensor is located in the control unit with display. Barometer enables to measure sea level pressure by setting of correction to altitude above sea level. maximum range of temperature measurement -30 to +105°C* probe including cable |
![]() |
||
T5410![]() |
temperature barometric pressure |
Thermometer-barometer outdoor and indoor use stem length 53mm maximum range of temperature measurement -30 to +80°C Barometer enables to measure sea level pressure by setting of correction to altitude above sea level. |
![]() |
||
T2414![]() |
barometric pressure |
Barometer Barometer enables to measure sea level pressure by setting of correction to altitude above sea level. |
![]() |


* Maximum temperature only at the measuring end with sensors. Relative humidity at temperature over +85°C is limited in accordance with the graph. Near plastic case with electronics maximum temperature is +80°C.

Figure: Working range of humidity probe
Figure:
T3419P humidity probe for compressed air up to 25 bars
- hard anodized duralumin, stainless steel mesh filter moulded in polycarbonate
Figure: Flow chamber SH-PP for compressed air measurement up to 25 bars - stainless steel
Included accessory:
Traceable calibration certificate from the manufacturer, instruction manual.
Calibration certificate with declared metrological traceability of etalons is based on requirements of EN ISO/IEC 17025 standard.
Example of calibration certificate.
TSensor - freeware for online configuring of the transmitter is ready to download.
SensorReader - freeware for online logging values from one transmitter to a PC disk file is ready to download. Program enables to alarm acoustically the PC user if adjusted alarm limits are exceeded. Recorded values in CSV format are easy to process in e.g. Excel. Program can be started several times on a computer at the same time to monitor values from several transmitters.
For online acquiring values from several transmitters, third party software is necessary to use. Tested functionality is with programs:
LabVIEW from National Instruments, by means of the free Modbus Library for LabVIEW.
For viewing of actual values and graphs from Tx41x transmitters following example of application was created in LabVIEW 8.5 :
RS232-485.vi
TIRS.NET - transmitters are also directly supported by program TIRS.NET from Coral s.r.o. company. Please download demo enabling full work with transmitters.
Tested functionality is also with programs:
ControlWeb,
EasyView.
OPTIONAL ACCESSORY | ||
---|---|---|
![]() |
SP021 | RS485/RS232 converter for for connection of Tx41x transmitters to serial COM port at the PC side, including ac/dc adapter. |
![]() |
TxxxxL |
Transmitter version with watertight male connector IP67 Lumberg RSFM4 instead of cable gland for easy connection/disconnection of the output. Specify please your order with letter L behind model code (e.g. TxxxxL). Minimum order of 10 pieces. |
![]() |
K1427 |
Female connector ELKA 4012PG7 for TxxxxL transmitters with male connector Lumberg for easy connection/disconnection of the output. Cable is easily connected to screw terminals of the connector. IP67 protection. |
![]() |
Transmitter version with blind lid without LCD. Specify please the requirement in your order. |
|
![]() |
F5200 | grey sensor cover with filter from stainless steel mesh, filtering ability 0,025mm |
![]() |
F5200B | black sensor cover with filter from stainless steel mesh, filtering ability 0,025mm for transmitter T3419, T7411 |
![]() |
F0000 | sintered bronze sensor cover. Filtering ability 0.025mm. |
![]() |
PP4 | flat plastic circular flange for duct mounting |
![]() |
PP90 | right-angled stain-less steel flange for wall mounting |
![]() |
SP004 | plastic gland for direct mounting of the humidity probe to a 29 mm diameter hole |
![]() |
MD036 | self adhesive Dual Lock for easy installation |
![]() |
SP006 | Tool for easy wire connection to terminals |
![]() |
F8000 | Solar radiation shield for transmitters T3419, T7411 with T+RH probe on a cable |
![]() |
SH-PP |
Flow chamber for compressed air measurement up to 25 bars - stainless steel DIN 1.4301
inlet and outlet connection - G1/8 thread humidity probe connection - G1/2 thread screw-coupling not included |
![]() |
A1510 | ac/dc adapter 100 to 240Vac/12Vdc |
![]() |
accessory for calibration and adjustment of humidity transmitter by the user |
Ask more information on availability of other communication protocols.
Transmitters are directly compatible with sixteen channel Comet data acquisition system MS.